Sterilisation of Meddins dental instruments

Instrument care

Sterilization

Sterilisation of dental instruments: step by step

This is a short guide for quick reference. For detailed information, please refer to the full version of the instructions.

Download the full version

7 steps

Processing procedure

Seven steps — from receiving the instrument to storage. The order and settings are the same as in the full instructions.

1

Delivery and inspection

  • Instruments are supplied non-sterile. Before use they must be thoroughly cleaned, rinsed and sterilised.
  • Check cleanliness and materials. Instruments with corrosion spots must not be used.
  • Return defective instruments to the manufacturer.
2

Pre-sterilisation cleaning

  • Performed in specially equipped areas.
  • After use, disinfect the instruments with suitable solutions, following the instructions for concentration, duration and temperature.
  • Avoid solutions containing NaCl — they cause corrosion.
  • Change cleaning solutions daily.
  • Do not wash instruments made of different metals together.
  • Instruments must be open during cleaning.
  • Avoid overloading trays with instruments.
  • Final rinse temperature: 80–90 °C; use demineralised water.
  • Instruments must be dry before packaging.
3

Sterilisation

  • Autoclave sterilisation: do not exceed 177 °C. Steam-sterilise for 4 minutes at 132 °C or 30 minutes at 121 °C. A 30-minute drying cycle is recommended.
  • The indicator tape on the packaging must change colour.
  • Dry-heat sterilisation: temperatures above 177 °C reduce the strength of the instruments and may cause darkening and corrosion.
4

Storage

  • Store sterile instruments in a dry, clean place at 5–40 °C, away from chemicals.
5

Use and preparation

  • Clean instruments immediately after use (within 30 minutes): remove coarse contamination and place them in a disinfection bath.
6

Disinfection

  • Manual disinfection: use a combined cleaning and disinfecting agent and follow the correct concentration, exposure time and temperature.
  • Ultrasonic disinfection: minimum cleaning time is 16 minutes. Use a pH-neutral enzymatic cleaner and rinse thoroughly with demineralised water.
  • Mechanical / automatic washer: use pH-neutral agents and follow the manufacturer’s instructions.
  • Thermal disinfector: use pH-neutral agents and follow the manufacturer’s instructions.
7

Drying and inspection

  • Instruments must be dry; check cleanliness and materials. If instruments are still contaminated, repeat the cleaning.
!

Attention!

Please note that aluminium instruments must not be treated with the following substances:

  • acids and alkalis
  • hydrogen peroxide solutions (including 6 %)
  • chlorine-containing agents
  • concentrated saline solutions
  • other aggressive or corrosive chemicals

Aluminium instruments

Safe pH for care

When choosing disinfecting and cleaning agents for aluminium instruments, consider the pH level: an unsuitable pH can damage the products.

Safe

pH 4 – 8.5

Safe range. Agents in this range are compatible with aluminium.

Caution

pH 8.5

Borderline level: gradual destruction of the anodised layer, especially with prolonged contact.

Dangerous

pH > 9.5

Dangerous level — may cause discolouration, loss of shine, a matte surface and crystallisation on the surface.

  • Use only agents with a neutral or slightly acidic / alkaline pH between 4 and 8.5 that are compatible with aluminium.
  • Do not use abrasive or metal objects for cleaning.
  • To preserve the quality and durability of the instruments, use only specialised disinfectants compatible with aluminium.